No. | Item Name | Unit of measurement | Requirements |
1 | Rated power supply of device |
| 220 V, 50 Hz |
2 | Power consumption does not exceed | W | 50 |
3 | Ramp time to operational mode after device has been switched on does not exceed | min. | 6 |
4 | Pulse energy is controllable within the range of | J | 0.05-1.0 |
5 | Voltage output pulse amplitude up to | kV | 9 |
6 | Pulse Repetition Modes: |
| -single pulse; -multiple pulse. |
7 | Number of pulses generated in an isolated batch | pcs | 2-99 |
8 | Pulse repetition frequency per burst is controllable from | Hz | 0.5-5 |
9 | Device allows continuous work in stand-by mode (with the Power switch on) for at least | h/days | 8 |
10 | Indicator that the device has a connection to earth |
| available |
11 | Lighted indication of the output pulse parameters |
| available |
12 | Flexible probe diameter | Fr | from 2.0 to 8.0 |
13 | Extra function: automated monitoring of the probe operational life and effectiveness |
| available |
The company "Medline" produces the nine types of probe – «Probe 2,0 Fr», «Probe 2,4 Fr», «Probe 2,7 Fr», « Probe 3,0 Fr», «Probe 3,3 Fr», « Probe 3,6 Fr», «Probe 4,5 Fr», «Probe 6,0 Fr», «Probe 8,0 Fr».
- Probes with tip diameter of 2,0 Fr (0.7 mm), 2,4 Fr (0.8 mm), 2,7 Fr (0.9 mm), 3,0 Fr (1.0 mm), 3,3 Fr (1.1 mm), 3.6 Fr (1.2 mm) length from 700 mm to 1250 mm are used for fragmentation of stones in the ureters and the kidney.
- Probes with tip diameter of 4,5 Fr (1.5 mm), 6,0 Fr (2.0 mm), 8,0 Fr (2.7 mm) length 700 mm are used for fragmentation of stones in the ureters and urinary bladder.
| |